CTDSMotor threshold · Hand and leg
Motor threshold assessment
Hand or leg hotspot assessment, with a pulse-by-pulse response counter and clinician-reviewed threshold record.
Earlier assessment context · review before reuse
Protocol: Not recorded · Scalp selection:
Earlier entries are context only. Confirm the selected protocol, actual coil, muscle, side and current assessment before use. Changes in earlier steps require review of later entries.
1. Prepare
Setup locks once trials are recorded. Reset before changing muscle, side, hotspot, coil or waveform.
3. Count responses
Lower failing bound: —% · Upper qualifying bound: —%
How this counting workflow works
- Start at 35% MSO only if appropriate for the patient and within the clinician-set ceiling. Adjust the starting value if necessary.
- Record Yes or No for each valid pulse, up to 10. Five positives in total qualify a search block, whether consecutive or not. Six negatives mean it cannot meet 5 of 10.
- Save the block to see the next suggested level: 5 points up after a failing level, or 5 points down after a qualifying level, until both bounds are found.
- Then refine downward from the upper bound in 1-point steps. Complete all 10 pulses at each of the final adjacent bounds before clinician verification.
This is a requested CTDS adaptation of relative-frequency counting. Groppa describes ascending to consistent EMG responses, then descending by 1%; the early-stop bracketing here is not that exact algorithm or a validated adaptive estimator. Buttons record choices; they do not control the stimulator.
Positive means: A valid resting MEP ≥50 µV peak-to-peak on EMG.
- Complete preparation checks
- Confirm the measured hotspot
- Enter the exact coil
- Enter the stimulator
- Choose the waveform
- Record coil orientation
- Check intensity and output ceiling
Enter a valid intensity and clinician-set output ceiling before recording.
| Block | % MSO | Yes / tested | Response sequence | Status |
|---|
4. Review the result
MT assessment report
MOTOR THRESHOLD — CTDS BRACKETING RECORD Response method: EMG. Positive = resting MEP ≥50 µV peak-to-peak; Yes/No entered by clinician. right hand; muscle: APB Device: not recorded; coil: not recorded; waveform: not recorded Hotspot relative to Cz: lateral -5 cm (negative=patient left), anterior 0.5 cm (negative=posterior). Manually entered schematic record, not image-derived anatomy. Hotspot verified: no; pose: not recorded Clinician-set ceiling: not recorded% MSO. Unfinished block: 0/10 at 35% MSO; responses: THRESHOLD NOT ESTABLISHED. Enter a valid intensity and clinician-set output ceiling before recording. CTDS custom 5-point bracketing / 1-point refinement; not the exact Groppa algorithm or a validated adaptive estimator. Final adjacent bounds require complete 10-pulse blocks. Groppa et al. 2012 doi:10.1016/j.clinph.2012.01.010. No hardware control or treatment-dose prescription.
Hand and leg thresholds, and thresholds measured with different coils, waveforms or response methods, are not interchangeable.
Method, source and limitations
Groppa et al. (2012), A practical guide to diagnostic transcranial magnetic stimulation: Report of an IFCN committee. Clinical Neurophysiology 123:858–882. §3.3.1 describes relative-frequency EMG threshold assessment; §6.1 and §9–10 inform hotspot, hand and leg guidance. CTDS adds the clinician-selected 5-point bracketing and early search prompts; these have not been clinically validated. Five consecutive positives are simply five total positives, not a separate threshold rule.
Read Groppa et al. ↗Visual observations do not establish EMG RMT. Visual threshold assessment can overestimate EMG threshold; confirm that the exact therapeutic device protocol permits the selected method.
Visual versus EMG threshold study ↗